The Dalat Nuclear Research Reactor (DNRR) was used to test and confirm the k0 factors needed for the k0-standardized neutron activation analysis (k0-NAA) Experimental validation of the k0 factors required by the k0-standardized neutron activation analysis (k0-NAA) using the Dalat Nuclear Research Reactor (DNRR) was done. The new k0 factors of 7 short and medium-lived radionuclides, were evaluated, including: 66Cu, 52V, 38Cl, 134mCs, 128I, 140La, and 56Mn, which were determined at DNRR. Perform the k0-NAA using the new k0 dataset to determine the concentrations of the elements in the reference materials (RMs) against the certified concentration values. The difference between the experimental results of determining the new k0 factors and the reference values (literature) is less than 7%. From there, applying the new k0 factors to k0-NAA at DNRR resulted in the elemental concentrations in the RMs compared to the certified values in the range of 7-11%. Exception for 134mCs, which were biased by 19%. This work conducted the experiments at DNRR, including the characterization of the Channel 7-1; calibration of the gamma-ray spectrometer using the HPGe detector; and the data processing that is required by the k0-NAA method with experimental parameters (neutron field parameters, detector efficiency, corrections, etc.). The "self-characterization" method for Channel 7-1 using synthetic multi-element standard samples (SMELS I) was the first use regarded as simple and convenient experimentally, with results consistent with the conventional method using independent monitors.
To sum up the result of the experimental validation of nuclear data in k0-NAA as using the Dalat nuclear research reactor (DNRR), it can be seen that Table 12 presents the experimentally determined k0 factor of 11 gamma-rays from 7 radionuclides of both short- and medium-lived nuclides using the DNRR. The results are compared with the recommended literature value and k0 reference. It was found that the k0 factors for the short-lived radionuclides 52V at 1434.1 keV and 66Cu at 1039.2 keV, as well as the medium-lived radionuclide 128I, are similar to the k0 factors for short-lived radionuclides 52V at the energy of 1434.1 keV and 66Cu at the energy of 1039.2 keV, as well as medium-lived radionuclides 128I, are determined and are in good agreement with the result of the k0 reference. It is important to note that 52V, 66Cu, and 128I emit only a single gamma ray, and there is no true coincidence effect. The ratio of the experiment to reference is 1.004 for both isotopes. 38Cl and 56Mn, which emit two gamma-rays with a high emission probability, the experimental results for determining the k0 factor are also in good agreement with the k0 reference. The differences between the experiment and reference values are less than 3%. For other radionuclides, such as 140La, which had three different gamma-ray energies, the k0 factor was generally in good agreement compared to the reference information. However, for 134mCs, a medium-lived radionuclide with an energy of 127.5 keV, the experiment value was compared to the k0 reference, and the Exp./Ref ratio. was 8.31E-01. The k0 factor of 134mCs has a high difference with the k0 reference value because this energy is in a low energy region (< 200 keV), and some effects from Compton or low X-ray may influence the results. Thus, 134mCs appear to be not good due to low uncertainty. The determination of the k0 factor measured the absolute uncertainty and was calculated based on the component of parameter error propagation. The uncertainties obtained uncertainties for determining the k0 factor for short- and medium-lived radionuclides were reported in the study.
                
              
                                            
                                
            
 
            
                 31 trang
31 trang | 
Chia sẻ: Kim Linh 2 | Ngày: 11/11/2024 | Lượt xem: 332 | Lượt tải: 0 
              
            Bạn đang xem trước 20 trang tài liệu Luận án Experimental validation of nuclear data in the k₀-standardized neutron activation analysis using the dalat research reactor, để xem tài liệu hoàn chỉnh bạn click vào nút DOWNLOAD ở trên
 lot of neutrons, samples can be neutron activated quickly and effectively [7]. 
3.2.2 The k0-NAA in Dalat nuclear research reactor 
The k0-NAA method has been implemented in the NAA laboratory at DNRR, 
and advancements in computer technology, evaluation programs, and laboratory 
instrument have led to re-validation of this standardization technique in laboratory. 
3.3 Practical Experiment in Dalat nuclear research reactor 
3.2.1 Sample preparation 
Table 3.2 Sample preparation. 
No. Name Type of sample Sample ID Mass (g) 
1 IRMM 530R Al-0.1%Au Comparator A0289 0.00411 
2 SMELS I Reference Material SM21a 0.03185 
3.2.2 Prepare monitors 
Table 3.2. Description of sample preparation conditions under the k0-IAEA program. 
No 
Standard 
sample 
Code 
Matrix 
sample 
Geometry 
Weight of 
sample (mg) 
1 Al-0.1%Au Au1102 Al-metal Foil 5.510 
2 None BLK-1 None None 4000 
3 NIST-2711a MO195 Plant leaves Powder/liquid 100.880 
 Page 19 of 29 
4 NIS-1570a SP4 Al-metal Foil 204.200 
5 NIST-1566b OT21 Plant leaves Powder/liquid 198.800 
6 Al-0.1%Au Au1100 Al-metal Foil/wire 7.200 
3.2.3 Measurement of monitors 
The samples were measured on a digital signal processing gamma 
spectrometer system comprising a 30% efficient HPGe detector (GMX 30190) and a 
transistor erasing preamplifier with high count rates, fast changes, and multiple 
functions. 
3.2.4 The use of specialized program k0-IAEA 
To use the k0-NAA method, you need to do two main things: find out the 
neutron spectral paramters at the sample irradiation position and check how well the 
gamma spectrometer system records. 
Table 3.3. Neutron spectral parameters of sample in channel 7-1 in the thermal column. 
Parameter Value Uncertainty 
th 3.80E+16 n.cm
-2s-1 1.524E+14 n.cm-2s-1 
f 9.86 0.10 
 -0.0676 0.0006 
3.2.5 Sample Irradiation of the seven short-lived radionuclides 
Table 3.5. The irradiation, decay and counting times for determination of k0 factor. 
Radionuclides Ti(s) Td (s) Tc (s) 
Distance to 
Detector (mm) 
66Cu, 52V 
70 s (~30 mg) 
5-10 m 300 51 
198Au, 38Cl, 
134mCs, 128I, 
140La, 56Mn 
1-2 h 900 51 
Table 3.6. SRMs and comparator information. 
Sample name Element 
Assigned Value 
± U mg/kg-1 
SMELS Type I 
Au 82.7 ± 1.7 
Cl 4330 ± 170 
Cs 897 ± 37 
Cu 3930 ± 120 
I 152 ± 5 
La 265 ± 10 
Mn 113.9 ± 3.3 
V 39 ± 1.6 
IRMM 530R Al-0.1%Au Au 1003 ± 12 
3.2.6 Calibration of neutron spectra in Channel 7-1 
Table 3.7. Irradiation channel neutron spectral parameter. 
Parameter Value Uncertainty 
th (n/m2/s) 3.80E+16 1.524E+15 
f 9.86 0.10 
 Page 20 of 29 
 -0.0676 0.0006 
Table 3.8. Nuclear parameters condition for k0-NAA. 
Element X (n, x) Y T1/2 Eγ (keV) 
Au 197Au (n, γ) 198Au 2.695 d 411.8 
Cl 37Cl (n, γ) 38Cl 37.24 min 1642.7, 2167.4 
Cs 133Cs (n, γ) 134mCs 2.903 h 127.5 
Cu 65Cu (n, γ) 66Cu 5.12 min 1039.2 
I 127I (n, γ) 128I 24.99 min 442.9 
La 139La (n, γ) 140La 1.678 d 328.8, 487.0, 596.2 
Mn 55Mn (n, γ) 56Mn 2.579 h 846.8, 1810.7 
V 51V (n, γ) 52V 3.75 min 1434.1 
3.2.7 Irradiation measurement 
The HPGe detector was used to measure gamma-rays emitted from the 
irradiated samples. The efficiency of absolute full energy peak detection was 
calibrated using a standard gamma-ray source of 152Eu, and the efficiency curve at 
51 mm to the detector. 
Chapter 4: Results and discussion 
4.1 Results 
The present study focused on determining seven radionuclides using the k0 
factors. The α and ƒ values obtained from Table 3.4 were used to calculate the k0 
factors by the equation (2.60) and k0 factors were determined using the "bare monitor" 
method. It was observed that the α and ƒ values obtained in this study agreed with the 
literature values of DNRR [24]. The k0 factors were determined using the "bare 
monitor" method. Table 4.1 presents the experimental results of the k0 factor 
determination for short- and medium-lived radionuclides using DNRR, and the 
obtained values were compared with the k0 references [16]. Regarding the 
contribution of the parameters such as sample mass, irradiation time, decay time, and 
counting time, are less than 0.1%, Q0 and f are about 1.0%, α is less than 1.5% and 
overall, they are about 3.24%, that all of which are in the agreement with the k0 
reference as shown in Table 4.3. 
Table 4.1. Experimental k0, Au -factors of the short- and medium-lived radionuclides 
compared to the k0 reference using Channel 7-1 facility at DNRR. 
Nuclides E𝛾 (keV) k0-experiment Uncertainty 
k0-Reference 
[16] 
Exp / Ref 
198Au 411.8 - - 1 - 
52V 1434.1 1.969E-01 1.237E-02 1.96E-01 1.004 
66Cu 1039.2 1.868E-03 1.156E-04 1.86E-03 1.004 
38Cl 1642.7 1.893E-03 1.197E-04 1.97E-03 0.961 
38Cl 2167.4 2.595E-03 1.635E-04 2.66E-03 0.976 
134mCs 127.5 4.768E-03 3.049E-04 5.74E-03 0.831 
128I 442.9 1.185E-02 7.582E-04 1.12E-02 1.065 
140La 328.8 2.687E-02 1.679E-03 2.87E-02 0.936 
140La 487 6.080E-02 3.789E-03 6.37E-02 0.955 
140La 1596.2 1.279E-01 8.013E-03 1.34E-01 0.955 
56Mn 846.8 4.887E-01 3.028E-02 4.96E-01 0.985 
56Mn 1810.7 1.314E-01 8.120E-03 1.35E-01 0.973 
Table 4.2. Relative standard uncertainties of parameters in the k0-NAA standardization. 
 Page 21 of 29 
Parameter Contribution 
Mass ~0.1% 
Q0 ~1.0% 
α ~1.5% 
f ~1.0% 
Efficiency ~2.0% 
Concentration ~1.5% 
Irradiation time, Decay time, Counting time Less than ~0.1% 
Overall ~3.24% 
4.2 Discussions 
4.2.1 Research results 
This section provides a detailed discussion of the experimental k0 factors 
obtained in this study. It is divided into three main parts: characterization of the 
neutron irradiation, calibration of gamma-ray spectrometers, and validation of the 
results using SRMs. Tables and figures are included in this section to support the 
discussion, and a summary is provided at the end. 
This study used seven short- and medium-lived radionuclides for neutron 
activation analysis. These samples were prepared in polyethylene boxes and 
irradiated at channel 7-1 of the DNRR for 70 seconds. The FWHM was 1.9 keV at 
1132 keV, and the efficiency was calibrated using 152Eu at 51 mm. The HPGe detector 
(GMX-30190) was used to count the samples with a relative efficiency of 30%. Decay 
times ranged from 5-10 minutes for short-lived radionuclides and 1-2 hours for 
medium-lived radionuclides, and the counting time was set at 300 seconds and 900 
seconds. All parameters are shown in Tables 3.8, Table 3.9, and Table 3.10. The 
accuracy of the reference materials was analyzed with the sample, and the nuclear 
reactions necessary for obtaining the results were characterized, particularly the 
short-lived irradiation of the SMELS materials. Gamma-ray spectrometers were 
calibrated using an HPGe detector (GMX-30190) to calibrate for energy, peak width, 
and full energy detection efficiency, as shown in Fig. 21. The calibration results 
showed that the value of 𝜙th is 3.80×106 n/m-2.s-1, ƒ is 9.86, and α is -0.0676. The 
calibration of the detector's energy efficiency detection and full peak energy was in 
good agreement with the standard value of the reference. The accuracy of the detector 
response functions for low activity was also calibrated. This allowed for 
simultaneously determining several other short-lived or detectable radionuclides 
contained elsewhere. However, it is important to consider the setting and self-
grinding of the matrix in the container, and the matrix was kept compact. The volume 
in the container is fixed to ensure stability and non-alteration of the matrix in long-
term usage. The ready standard has also fixed the volume that cannot be changed 
during transportation and usage of the standard, and natural matrices were used in the 
properties as close to the analyzed sample as possible. 
 Page 22 of 29 
Table 4.3. Values of k0,Au determined by this work and the other authors. 
Radionuclides 
(Energy, keV) 
This work 
± Unc. 
Reference [16] 
± Unc. 
Reference [26] 
± Unc. 
IAEA [41] 
± Unc. 
Reference [42] 
± Unc. 
52V (1434.1) 1.97E-1 ± 1.24E-2 1.96E-1 ± 2.35E-3 2.00E-1 ± 2.40E-3 1.96E-1 ± 2.35E-3 2.00E-1 ± 2.20E-3 
66Cu (1039.2) 1.87E-3 ± 1.16E-4 1.86E-3 ± 9.30E-6 1.76E-3 ± 8.80E-6 1.86E-3 ± 9.30E-6 2.03E-3 ± 2.03E-5 
38Cl (1642.7) 1.89E-3 ± 1.2E-4 1.97E-3 ± 2.76E-5 2.30E-3 ± 3.22E-5 1.97E-3 ± 2.76E-5 2.06E-3 ± 2.88E-5 
38Cl (2167.4) 2.60E-3 ± 1.64E-4 2.66E-3 ± 3.46E-5 2.72E-3 ± 3.81E-5 2.66E-3 ± 3.46E-5 2.75E-3 ± 3.58E-5 
134mCs (127.5) 4.77E-3 ± 3.05E-4 5.74E-3 ± 9.76E-5 5.35E-3 ± 9.10E-5 5.74E-3 ± 6.89E-5 - 
128I (442.9) 1.19E-2 ± 7.59E-4 1.12E-2 ± 1.90E-4 1.79E-2 ± 3.04E-4 1.16E-2 ± 1.62E-4 - 
140La (328.8) 2.69E-2 ± 1.68E-3 2.87E-2 ± 2.87E-4 2.48E-2 ± 2.48E-4 2.87E-2 ± 2.87E-4 2.83E-2 ± 3.11E-4 
140La (487) 6.08E-2 ± 3.79E-3 6.37E-2 ± 5.73E-4 5.77E-2 ± 5.19E-4 6.37E-2 ± 5.73E-4 6.26E-2 ± 6.26E-5 
140La (1596.2) 1.28E-1 ± 8.02E-3 1.34E-1 ± 1.47E-3 1.28E-1 ± 1.41E-3 1.34E-1 ± 1.47E-3 1.30E-1 ± 5.20E-4 
56Mn (846.8) 4.89E-1 ± 3.03E-2 4.96E-1 ± 2.98E-3 5.01E-1 ± 3.01E-3 4.96E-1 ± 2.98E-3 5.02E-1 ± 5.02E-3 
56Mn (1810.7) 1.31E-1 ± 8.13E-3 1.35E-1 ± 5.40E-4 1.38E-1 ± 5.52E-4 1.35E-1 ± 5.40E-4 1.37E-1 ± 1.23E-3 
 * Remark: Unc. = Absolute uncertainty.
 Page 23 of 29 
Table 4.4. The comparison of analysis results (mg/kg) based on the experimented 
and referenced k0,Au factors. 
Elements 
Radio- 
nuclides 
Ce Ue Cr Ur Cc Uc cC/ eC cC/ rC 
Using k0,Au factors 
(experiment) 
Using k0,Au 
factors 
(reference) 
Certified values 
Experiment/ 
Certified 
Reference/ 
Certified 
V 52V 38 2 42 2 39 2 0.97 1.08 
Cu 66Cu 4284 171 4039 7 3930 118 1.09 1.03 
Cl 38Cl 4407 176 4313 173 4330 173 1.02 1.00 
Cs 134mCs 934 56 776 47 897 36 1.04 0.87 
I 128I 151 8 143 7 152 5 0.99 0.94 
La 140La 267 19 254 18 265 11 1.01 0.96 
Mn 56Mn 111 3 108 3 114 3 0.97 0.95 
Remarks: 
 Ce and Ue - The experimental concentrations and uncertainties, respectively 
obtained using the k0,Au values in this work; 
 Cr and Ue - The reference concentrations and uncertainties, respectively obtained 
using the k0,Au values from references of DeCort, 2003 [17]; 
 Cc and Uc - The certified concentrations and uncertainties of the certified reference 
material (SMELS Type-I). 
The comparison of the determined k0,Au factors (Ce) experimentally with those 
from references (Cr) showed a good agreement with certified values (Cc) for various 
elements (see Table 4.4). The ratios Ce/Cc and Cr/Cc, which measure agreement, are 
consistently close to one, confirming that both the experimentally determined and 
reference k0,Au factors are reliable for analysis of the elemental concentrations of 
interest. We carefully considered uncertainty, represented by Ue and Ur for the 
experimental and reference k0,Au factors resulted in a high accuracy of the k0-based 
method as demonstrated in this study. However, upon closer investigation, there are 
slight differences, especially for Cu (this work) and Cs (reference). In these cases, 
experimentally determined and reference values deviated from certified values of 9% 
and 13%, respectively. This observation suggests a further research to the specific 
conditions or factors affecting the determination of k0,Au factors for these elements. 
This detailed investigation would be crucial for improving the method and ensuring 
higher accuracy for analyses in the future. In summary, the concentrations and 
uncertainties of the experimental and reference of k0,Au factors. The value of k0,Au 
factors obtained in this work for some short and medium-lived radionuclides (52V, 
66Cu, 38Cl, 134mCs, 128I, 140La, and 56Mn) have been assessed by analyzing the 
SMELS Type-I. The deviation of the results of analysis for SMELS Type-I with the 
experiment and reference k0,Au factors as compared with the certified values were 
within ± 9.0% and ± 13.0%, respectively. The comparative results showed that the 
experiment k0,Au factors done by this study were acceptable and better agreement 
as compared with the reference ones. 
 Page 24 of 29 
4.3 Conclusions and future work 
4.3.1 Conclusions 
The Dalat Nuclear Research Reactor (DNRR) was used to test and confirm the 
k0 factors needed for the k0-standardized neutron activation analysis (k0-NAA) 
Experimental validation of the k0 factors required by the k0-standardized neutron 
activation analysis (k0-NAA) using the Dalat Nuclear Research Reactor (DNRR) was 
done. The new k0 factors of 7 short and medium-lived radionuclides, were evaluated, 
including: 66Cu, 52V, 38Cl, 134mCs, 128I, 140La, and 56Mn, which were determined at 
DNRR. Perform the k0-NAA using the new k0 dataset to determine the concentrations 
of the elements in the reference materials (RMs) against the certified concentration 
values. The difference between the experimental results of determining the new k0 
factors and the reference values (literature) is less than 7%. From there, applying the 
new k0 factors to k0-NAA at DNRR resulted in the elemental concentrations in the 
RMs compared to the certified values in the range of 7-11%. Exception for 134mCs, 
which were biased by 19%. This work conducted the experiments at DNRR, 
including the characterization of the Channel 7-1; calibration of the gamma-ray 
spectrometer using the HPGe detector; and the data processing that is required by the 
k0-NAA method with experimental parameters (neutron field parameters, detector 
efficiency, corrections, etc.). The "self-characterization" method for Channel 7-1 
using synthetic multi-element standard samples (SMELS I) was the first use regarded 
as simple and convenient experimentally, with results consistent with the 
conventional method using independent monitors. 
To sum up the result of the experimental validation of nuclear data in k0-NAA 
as using the Dalat nuclear research reactor (DNRR), it can be seen that Table 12 
presents the experimentally determined k0 factor of 11 gamma-rays from 7 
radionuclides of both short- and medium-lived nuclides using the DNRR. The results 
are compared with the recommended literature value and k0 reference. It was found 
that the k0 factors for the short-lived radionuclides 52V at 1434.1 keV and 66Cu at 
1039.2 keV, as well as the medium-lived radionuclide 128I, are similar to the k0 factors 
for short-lived radionuclides 52V at the energy of 1434.1 keV and 66Cu at the energy 
of 1039.2 keV, as well as medium-lived radionuclides 128I, are determined and are in 
good agreement with the result of the k0 reference. It is important to note that 52V, 
66Cu, and 128I emit only a single gamma ray, and there is no true coincidence effect. 
The ratio of the experiment to reference is 1.004 for both isotopes. 38Cl and 56Mn, 
which emit two gamma-rays with a high emission probability, the experimental 
results for determining the k0 factor are also in good agreement with the k0 reference. 
The differences between the experiment and reference values are less than 3%. For 
other radionuclides, such as 140La, which had three different gamma-ray energies, the 
k0 factor was generally in good agreement compared to the reference information. 
However, for 134mCs, a medium-lived radionuclide with an energy of 127.5 keV, the 
experiment value was compared to the k0 reference, and the Exp./Ref ratio. was 
8.31E-01. The k0 factor of 134mCs has a high difference with the k0 reference value 
because this energy is in a low energy region (< 200 keV), and some effects from 
Compton or low X-ray may influence the results. Thus, 134mCs appear to be not good 
due to low uncertainty. The determination of the k0 factor measured the absolute 
uncertainty and was calculated based on the component of parameter error 
propagation. The uncertainties obtained uncertainties for determining the k0 factor for 
short- and medium-lived radionuclides were reported in the study. 
 Page 25 of 29 
4.3.2 New points of discussion 
 To validate the k0 experiment used in my own experiments, I compare them 
with those reported by different authors in the field [16, 26, 41, 42]. This will involve 
a thorough review of the literature, looking for studies that have used similar reactor 
and neutron spectra to those used in my experiments. This comparison will be critical 
for validating the nuclear data used in my experiments, as well as for ensuring that 
my experimental results are in good agreement with the reference values. Any 
discrepancies or differences between my k0 values and those reported in the literature 
will need to be carefully investigated, to ensure that my results are as accurate as 
possible. In addition to comparing my k0 experiment with those reported in the 
literature, I will also need to consider any potential sources of error in my own 
experiments. This may involve a careful analysis of the data, looking for any 
systematic errors or biases that could affect the accuracy of my results. I will also 
need to consider the effects of any uncertainties in the nuclear data used in my 
experiments, and to quantify these uncertainties wherever possible. 
4.3.2 Future plan 
A Future plan in this research area is proposed to investigate and re-determine 
the k0 factors of very short-lived radionuclides (less than 20 seconds), such as 77mSe 
and 20F as follows: 
(1) There is a lack of experimental data for these radionuclides, and additional 
experiments are needed to improve their accuracy and reduce the uncertainty 
of their k0 values. 
(2) Re-determination of k0 factors for radionuclides with high uncertainties, such 
as 133Ba is needed. This will help to improve the accuracy of k0 factor and 
increase the reliability of the k0-NAA method for the element. 
(3) Determination of the k0 factor for 152Eu at low energy (46 keV) is proposed 
because it is missing in the current k0 dataset. 
Papers published using for the dissertation 
1. Ho Manh Dung, Tran Tuan Anh, Tran Quang Thien, Ho Van Doanh, Truong 
Truong Son, Phonesavanh Lathdavong, Analysis of automobile window glass 
samples by the k0-based Neutron Activation Analysis for forensic applications, 
Journal of Radioanalytical and Nuclear Chemistry, Volume 332 (2023) 3493-
3498; 
2. Ho Manh Dung, Phonesavanh Lathdavong, A review of nuclear data for the k0-
based neutron activation analysis, Journal of Nuclear Science and Technology, 
Vietnam, Vol.9, No. 1 (2019) 28-33; 
3. Phonesavanh Lathdavong, Ho Manh Dung, Tran Quang Thien, Doanh Van Ho, 
Son Truong Truong, Self-characterization of irradiation facility using synthetic 
multi-element standard (SMELS) for the determination of k0-factors of seven 
radionuclides of interest, Journal of Nuclear Science and Technology, Vietnam, 
accepted for publishing in September 2023; 
4. Ho Manh Dung, Tran Tuan Anh, Ho Van Doanh, Tran Quang Thien, Nguyen Thi 
Tho, Trinh Van Cuong, Truong Truong Son, Phonesavanh Lathdavong, 
Evaluation of “k0-DALAT” software for the k0-Standardized neutron activation 
analysis, VINANST-14, Da Lat, 09-10/12/2021. 
 Page 26 of 29 
References 
[1] Gibbons, D., & Kruger, P. (1971). Principles of activation analysis. John 
Wiley & Sons. Wiley-Interscience. 
[2] Hamidatou, L. A. (2019). Advances technologies and applications of neutron 
activation analysis. IntechOpen. https://www.intechopen.com/books/6336. 
DOI: 10.5772/intechopen.69725. 
[3] IAEA (2001). Use of research reactors for neutron activation analysis (IAEA-
TECDOC-1215). IAEA-Vienna. 
[4] Speakman, R. J., & Glascock, M. D. (2007). Acknowledging fifty years of 
neutron activation analysis in archaeology. Archaeometry, 49, 179-183. 
[5] Byrne, A. R., & Kucera, J. (1997). Role of the self-validation principle of NAA 
in the quality assurance of bioenvironmental studies and in the certification of 
reference materials. Proc. Int. Symp. Harmonisation of Health-related 
Environ. Measurements using Nucl. and Isotopic techniques (Hyderabad, 4–7 
Nov. 1996), IAEA Vienna (1997) 223–238. 
[6] k0-International Scientific Committee (k0-ISC). Classic k0 Database. 2012. 
[7] Dung, H. M., & Hien, P. D. (2003). The application and development of k0-
standardization method of neutron activation analysis at Dalat research 
reactor. Journal of Radioanalytical and Nuclear Chemistry, 257(3), 643-647. 
[8] Greenberg, R. R., Bode, P., & Fernandes, E. A. D. N. (2011). Neutron 
activation analysis: A primary method of measurement. Spectrochimica Acta, 
B66, 193-241. 
[9] Kučera J., Byrne A.R.: Role of the self-validation principle of NAA in the 
quality assurance of bioenvironmental studies and in the certification of 
reference materials. IAEA-SM-344/8, IAEA Vienna, 1997. 
[10] De Corte, F. (1999). Thirty years of NAA developments and applications at 
the Thetis reactor of the Institute for Nuclear Sciences, Gent. Czechoslovak 
journal of physics, 49, 247-254. 
[11] Freitas, M. C., Reis, M. A., Marques, A. P., Almeida, S. M., Farinha, M. M., 
De Oliveira, O. M., Ventura, M. G., Pacheco, A. M., & Barros, L. P. (2003). 
Monitoring of environmental contaminants: 10 years of application of k0-
INAA. Journal of Radioanalytical and Nuclear Chemistry, 257(3), 621-625. 
[12] Byrne, A. R. (1993). Review of neutron activation analysis in the 
standardization and study of reference materials, including its application to 
radionuclide reference materials. Fresenius' journal of analytical chemistry, 
345, 144-151. 
[13] Řanda, Z., Kučera, J., & Soukal, L. (2003). Elemental characterization of the 
new Czech meteorite ‘Moravka’ by neutron and photon activation analysis. 
Journal of Radioanalytical and Nuclear Chemistry, 257, 275-283. 
[14] Santos J. O., Munita C. S., Toyota R. G., Vergne C., Silva R. S., Oliveira P. 
M. S.: The archaeometry study of the chemical and mineral composition of 
pottery from Brazil’s Northeast. J. Radioanal. Nucl. Chem. 281 (2009), p. 
189–192. 
[15] Pomme, S., Hardeman, F., Robouch, P., Etxebarria, N., & Arana, G. (1997). 
Neutron activation analysis with k0-standardisation: general formalism and 
procedure. Applied Radiation and Isotopes, 48, 1275-1288. 
 Page 27 of 29 
[16] De Corte, F., & Simonits, A. (2003). Recommended nuclear data for use in the 
k0 standardization of neutron activation analysis. Atomic Data and Nuclear 
Data Tables, 85, 47-65. 
[17] De Corte, F. D. (1987). The k0-standardization method - a move to the 
optimization of NAA. Habilitation Thesis, University of Gent. 
[18] De Corte F., De Wispelaere A. (2005), "The use of a Zr-Au-Lu alloy for 
calibrating the irradiation facility in k0-NAA for general neutron spectrum 
monitoring", J. Radioanal. Nucl. Chem. 263, pp.653-657. 
[19] Eguskiza, M., Robouch, P., Wätjen, U., & De Corte, F. (2003). The 
preparation and homogeneity study of synthetic multi-element standards 
(SMELS) for QC/QA of k0-NAA. Journal of Radioanalytical and Nuclear 
Chemistry, 257, 669-676. 
[20] Vermaercke, P., Robouch, P., Eguskiza, M., De Corte, F., Kennedy, G., 
Smodiš, B., ... Lin, X. (2006). Characterisation of synthetic multi-element 
standards (SMELS) used for the validation of k0-NAA. Nuclear Instruments 
and Methods in Physics Research Section A: Accelerators, Spectrometers, 
Detectors and Associated Equipment, 564, 675-682. 
[21] Dung, H. M. (2016). Quality evaluation of the k0-standardized neutron 
activation analysis at the Dalat research reactor. Journal of Radioanalytical and 
Nuclear Chemistry, 309(1), 135-143. 
[22] Doanh, H. V., & Dung, H. M. (2018). The upgrading of the cyclic neutron 
activation analysis facility at the Dalat research reactor. Journal of 
Radioanalytical and Nuclear Chemistry, 315, 301-308. 
[23] Ho Manh Dung, Tran Quang Thien, Ho Van Doanh, Truong Truong Son, 
Phonesavanh Lathdavong, Development of a PC program for the k0‑based 
epithermal neutron activation analysis, Journal of Radioanalytical and Nuclear 
Chemistry, Volume 332 (2023) 3469–3474. 
[24] Vu, C. D., Thien, T., Doanh, H., Quyet, P., Anh, T., & Dien, N. (2014). 
Characterization of neutron spectrum parameters at irradiation channels for 
neutron activation analysis after full conversion of the Dalat nuclear research 
reactor to low enriched uranium fuel. Journal of Nuclear Science and 
Technology, Vol. 4(1), 70-75. 
[25] Farina Arboccò, F., Vermaercke, P., Smits, K., Sneyers, L., & Strijckmans, K. 
(2014). Experimental determination of k0, Q0 factors, effective resonance 
energies and neutron cross-sections for 37 isotopes of interest in NAA. Journal 
of Radioanalytical and Nuclear Chemistry, 302(1), 655-672. 
[26] De Corte, F., & Simonits, A. (1989). k0-Measurements and related nuclear 
data compilation for (n, γ) reactor neutron activation analysis: IIIb: tabulation. 
Journal of Radioanalytical and Nuclear Chemistry, 133, 43-130. 
[27] Arbocco, F. F., Vermaercke, P., Smits, K., Sneyers, L., Strijckmans, K. (2013). 
Experimental determination of k0, Q0, Er factors and neutron cross-sections 
for 41 isotopes of interest in Neutron Activation Analysis. J. Radioanal. Nucl. 
Chem., 296, 931-938. 
[28] Blaauw, M. (1994). The holistic analysis of gamma-ray spectra in instrumental 
neutron activation analysis. Nuclear Instruments and Methods in Physics 
Research Section A: Accelerators, Spectrometers, Detectors and Associated 
Equipment, 353(1-3), 269-271. 
 Page 28 of 29 
[29] Kučera, J., Kamenik, J., & Povinec, P. P. (2017). Radiochemical separation of 
mostly short-lived neutron activation products. Journal of Radioanalytical and 
Nuclear Chemistry, 311, 1299-1307. 
[30] Kučera, J., & Zeisler, R. (2004). Do we need radiochemical separation in 
activation analysis? Journal of radioanalytical and nuclear chemistry, 262(1), 
255-260. 
[31] Kučera J., Řanda Z.: The present role of neutron and photon activation analysis 
in determination of trace elements. 3rd, 85 Symp. on Nuclear Chemistry, 
Halifax, NS, Canada, 11–14 June, 2001. 
[32] Use of research reactors for neutron activation analysis. IAEA-TECDOC-
1215, IAEA Vienna, 2001. 
[33] Currie, L. A. (1999). Detection and quantification limits: origins and historical 
overview. Analytica Chimica Acta, 391(2), 127-134. 
[34] Taskaeva, M., Taskaev, E., & Penev, I. (1996). On the preparation of 
efficiency calibration standards for gamma-ray spectrometers. Applied 
Radiation and Isotopes, 47, 981-990. 
[35] De Corte, F., Bellemans, F., De Neve, P., Simonits, A. (1994). The use of a 
modified Westcott-formalism in the ko-standardization of N The state of 
affairs. Journal of radioanalytical and nuclear chemistry, 179, 93-103. 
[36] De Corte, F., Simonits, A., De Wispelaere, A., & Hoste, J. (1987). Accuracy 
and applicability of the k0-standardization method. Journal of Radioanalytical 
and Nuclear Chemistry, 113(1), 145-161. 
[37] De Corte, F. et, al., (1972). In Book of Abstract, FOUR DECADES OF k0-
NAA: AN APPRAISAL, Neutron Activation Analysis, 17, 115. 
[38] De Corte, F., & Simonits, A. (1994). Vademecum for k0-users. DMS 
Research, Geleen. 
[39] Ellison, S. L., & Williams, A. (2012). Quantifying uncertainty in analytical 
measurement. 
[40] Truong Truong Son, Ho Van Doanh, Ho Manh Dung, Determination of k0 
factors of short-lived nuclides 46mSc and 110Ag for the k0-NAA, Journal of 
Nuclear Engineering and Technology, 55 (2023) 3202-3205. 
[41] IAEA (2020). The k0-database by k0-ISC 2020. 
[42] Farina Arboccò, F., Vermaercke, P., Smits, K., Sneyers, L., Strijckmans, K. 
(2014). Experimental determination of k0, Q0 factors, effective resonance 
energies and neutron cross-sections for 37 isotopes of interest in NAA. Journal 
of Radioanalytical and Nuclear Chemistry, 302, 655-672. 
[43] Acharya, R., Holzbecher, J., & Chatt, A. (2012). Determination of k0-factors 
of short-lived nuclides and application of k0-NAA to selected trace elements. 
Nuclear Instruments and Methods in Physics Research Section A: 
Accelerators, Spectrometers, Detectors and Associated Equipment, 680, 1-5. 
[44] D’Agostino, G., Di Luzio, M., & Oddone, M. (2018). An uncertainty 
spreadsheet for the k0-standardisation method in neutron activation analysis. 
Journal of Radioanalytical and Nuclear Chemistry, 318(2), 1261-1269. 
[45] Molnar, G. L., Revay, Z., & Szentmiklosi, L. (2003). New perspectives for 
very short-lived neutron activation analysis. Journal of Radioanalytical and 
Nuclear Chemistry, 255(2), 157-163. 
 Page 29 of 29 
[46] Acharya, R., Nair, A., Reddy, A., & Manohar, S. (2002). Validation of a 
neutron activation analysis method using k0-standardization. Applied 
Radiation and Isotopes, 57(3), 391-398. 
[47] Jaćimović, R., Stibilj, V. (2010). Determination of Q0 and k0 factors for 75Se 
and their validation using a known mass of Se on cellulose. Nuclear 
Instruments and Methods in Physics Research Section A: Accelerators, 
Spectrometers, Detectors and Associated Equipment, 622, 415-418. 
[48] Acharya, R., & Chatt, A. (2003). Characterization of the Dalhousie University 
SLOWPOKE-2 reactor for k0-NAA and application to medium-lived 
nuclides. Journal of Radioanalytical and Nuclear Chemistry, 257(3), 525-
529.2012 
[49] Acharya, R., Swain, K., & Reddy, A. (2010). Analysis of SMELS and 
reference materials for validation of the k0-based internal mono-standard 
NAA method using in-situ detection efficiency. Nuclear Instruments and 
Methods in Physics Research Section A: Accelerators, Spectrometers, 
Detectors and Associated Equipment, 622, 411-414. 
[50] Alghem, L., Ramdhane, M., Khaled, S., & Akhal, T. (2006). The development 
and application of k0-standardization method of neutron activation analysis at 
Es-Salam research reactor. Nuclear Instruments and Methods in Physics 
Research Section A: Accelerators, Spectrometers, Detectors and Associated 
Equipment, Vol.556, 386-390.